02 · Focus Area 2 · Modernising legacy visualisations

Read the control charts

Every HPLC run includes control injections: a reference sample with a known answer. If the control drifts, the samples measured alongside it are suspect. A control chart plots each control result in time order against limits set from the process's own variation.

The 2021 team rebuilt these charts with one reusable function: the mean, the sample standard deviation, bands at 1, 2 and 3σ, and a printed list of runs beyond 3σ. That method is the default here. The run rules (with their false-alarm cost), the I-MR option, the disturbance tool and the optional AI explanation are additions for this revival.

Choose a chart

Control assay

Cation exchange. Charge-variant profile: acidic, main and basic peaks.

Rel.Area Basic · individuals, μ ± 3σ

n = 151

Rel.Area Basic control chart for Control std B (system suitability), CEX assay, 151 injections. 2 beyond 3 sigma, 2 rule signals.8910111213UCL12.403+2σ+1σCL10.764−1σ−2σLCL9.1252019-072019-112020-052020-112021-042021-08WE1

Hover or focus the chart and use the arrow keys to read individual injections. Filled red points are beyond 3σ; ringed points belong to another active rule.

Centre line

10.764

Sigma (sample SD)

0.546

Beyond 3 sigma

2

2 signals from 1 active rule

Explain this chartoptional · AI · your own key

Sends only this chart's numeric summary (limits, rules, flagged points, run statistics and the published false-alarm figures) to Anthropic, from your browser. What is sent

Out of control

The notebook printed one line per point beyond the 3σ limits. Same format, current chart.
Out of control:
20200217_CEX_Vega_S3: 7.854
20200217_CEX_Vega_S3: 7.645

Sequences flagged across measurements

For Control std B (system suitability), every measurement charted the 2021 way. Streams come from joining each sequence's sample injections back to the parsed lab sheet on Sample Name, as the team did.
Sequences with points beyond 3 sigma, by number of measurements affected
Sequence NameMeasurements beyond 3σStreams in this run
20200217_CEX_Vega_S3
Upstream
20201223_CEX_Nova_S1
Formulation
20201116_CEX_Atlas_S3
Concentrate